Keratometry & Corneal Power Measurement
Overview
The cornea does most of the eye's focusing work before light ever reaches the implanted lens, and keratometry is how an IOL formula accounts for that. A keratometer doesn't measure corneal power directly. It measures corneal curvature, then converts that curvature into an estimated power using an assumed refractive index for the cornea. This conversion is reliable for a normal, unoperated cornea. It breaks down once the cornea's shape or layered structure has changed, most often after refractive surgery (see IOL power calculation in atypical eyes for the post-LASIK case specifically).
Keratometry Principles
A keratometer measures the size of a reflected image, a "mire," off the tear film covering the corneal surface. The cornea's front surface acts as a convex mirror, and the size of that reflection is directly related to the radius of curvature producing it. Standard keratometry captures only the anterior corneal surface. It does not measure the posterior surface at all; total corneal power is instead estimated from a fixed, assumed index of refraction, calibrated to represent the average relationship between anterior and posterior corneal curvature in a normal cornea.
That assumption is exactly what makes keratometry unreliable once a cornea's anterior-to-posterior curvature relationship no longer matches the population average it was calibrated against. The clearest case is a cornea that has had refractive surgery: the anterior surface is reshaped while the posterior surface is left alone, decoupling the two in a way the keratometer's built-in assumption never anticipated.
Keratometry Units
Keratometry values used in IOL power formulas are expressed in diopters (D), the same unit used for spectacle and IOL power, rather than a purely geometric unit like millimeters of radius of curvature. This is deliberate. Converting the measured curvature into diopters at the point of measurement lets the keratometry value plug directly into a power calculation formula alongside the eye's other refractive inputs, with no extra unit conversion step. A steeper cornea, meaning a smaller radius of curvature, reads as a higher diopter value; a flatter cornea reads lower.
White-to-White Corneal Diameter
White-to-white (WTW) is a separate biometric parameter from keratometry. It estimates the horizontal visible corneal diameter, limbus to limbus, and says nothing about corneal curvature or power. Its main clinical uses are sizing phakic IOLs, where the implant has to match the eye's anterior chamber dimensions to avoid angle crowding or an undersized lens that rotates or decenters, and assessing alignment for toric IOLs, where stable reference landmarks confirm the lens has rotated to the correct meridian.
Don't conflate white-to-white with keratometry or axial length. WTW is a linear diameter measurement used for lens sizing and toric alignment, not a power or curvature measurement, and it plays essentially no role in standard monofocal IOL power calculation for a routine phakic-to-pseudophakic cataract case.
Corneal Power's Role in IOL Formulas
Keratometry does more inside a modern IOL power formula than simply get added to a power equation alongside axial length. Third- and fourth-generation formulas also use it, often together with anterior chamber depth, to help predict effective lens position (ELP): the anteroposterior position the implanted IOL is expected to settle into inside the capsular bag after surgery. ELP has a substantial effect on the final refractive outcome, so an error in the keratometry value propagates twice, once directly through the power equation and once indirectly through a mis-predicted lens position.
The newest generation of formulas, among them Holladay 2, Barrett Universal II, and Olsen, refine ELP prediction further by adding measured lens thickness as another variable: an optical biometer measurement of the crystalline lens itself, on the reasoning that a thicker natural lens tends to correlate with where the replacement IOL will settle. Lens thickness isn't a keratometry value. It's captured by the same optical biometer scan and used alongside keratometry to sharpen the ELP estimate these newer formulas produce.
It's tempting to treat keratometry error as small and forgivable next to axial length. In a normal eye it usually is. But keratometry feeds both the direct power calculation and the effective lens position prediction, so an error from an atypical cornea, whether from prior refractive surgery, irregular astigmatism, or dry eye distorting the tear film at the moment of measurement, can produce a disproportionately large refractive surprise for what looks like a modest measurement error.
High-Yield Pearls
- Keratometry measures anterior corneal curvature, then estimates total corneal power using an assumed refractive index; it does not directly measure the posterior surface.
- Keratometry values are expressed in diopters so they plug directly into IOL power formulas without unit conversion.
- White-to-white is a diameter measurement for phakic IOL sizing and toric alignment, not a power measurement; don't confuse it with keratometry.
- Newer-generation formulas (Holladay 2, Barrett Universal II, Olsen) add measured lens thickness to keratometry-based effective lens position prediction.
- An unstable tear film at the moment of measurement can distort keratometry the same way it distorts topography; optimize the ocular surface before biometry, not after a surprising result.